AppendixA
Figure A.1 Profile amplitudes of bulk water, oil and emulsified water
A.4. Procedure for T2, restricted diffusion and profile measurements
Referto Instruction Manual of MARAN.
1. Perform tuning for 01, 90o and 180° rf pulses.
2. Load script file set_prarameters.ris to set default parameters for all the
measurement.
3. Load script file t2_diff_profile.ris, load file G1_emul for magnetic gradient g list
(G1 in script) and file OneD4_500ms (for one-time measurement) or
RepeatD4_500ms (for successive measurement as function of time) for
199